ZHEJIANG SHENGDI TECHNOLOGY INC.

Monofásico/instrumento trifásico de la prueba del metro de la energía

Inicio
Productos
Sobre nosotros
Visita a la fábrica
Control de Calidad
Contacto
Solicitar una cotización
Inicio ProductosWithstand Voltage Meter Test Equipment

HS-6121 Single Phase AC Withstand Voltage Test Bench

HS-6121 Single Phase AC Withstand Voltage Test Bench

HS-6121 Single Phase AC Withstand Voltage Test Bench

Datos del producto:

Place of Origin: Zhejiang
Nombre de la marca: Shengdi Electrical
Certificación: SGS,CNAS
Model Number: HS6121

Pago y Envío Términos:

Minimum Order Quantity: 1 set
Precio: USD$1~100000/SET
Packaging Details: Wooden Packing
Delivery Time: 35days
Payment Terms: T/T,L/C
Supply Ability: 50 Set/Year
Contactar ahora
Descripción detallada del producto

HS-6121 withstand voltage test bench for single-phase watt-hour meter

 

is new-generation withstand voltage test bench for watt-hour meters which is introduced to the market by Shengdi Electrical in a powerful and sensational way to go with the market demand. That test bench is designed with both combinable and splittable high-voltage outputting and testing assemblies and the test process can be performed both manually and automatically. With meet well current market demand for watt-hour meter’s withstand voltage test. So it is an ideal withstand voltage test bench for single-phase multi-station watt-hour meter.

 

 

Technical Parameters

 

WITHSTAND VOLTAGE TESTER PARAMETERS

1.Input voltage: AC 220V±10%

2.Output voltage:0KV-5KV AC/50Hz, Accuracy: ±5%

3.Output capacity:500VA

4.Current leakage: Maximum current leakage:100Ma

5.Voltage test operation modes: both manually and automatically

6.Time control: 0~99s optional, Accuracy: ±5%

 

 

TEST BENCH PARAMETERS

1.Number of meters to be hung: 10(standard),0-60 single-phase watt-hour meters optional

2.Max. power consumption:500VA

3.Current leakage per meter station:5mA±5%

4.Output waveform distortion:<5%

5.Insulation resistance of voltage loop, current loop, and all other circuits to ground: all larger than 10MΩ

6.Combined error of voltage output: not larger than±2%

7.Operating environment: Temperature: 0~45°,relative humidity: not larger than 85%

8.External form of test bench

   Meter hanger:2200×600×1920mm(48 meter stations), test bench:600×600×1200mm

 

Contacto
Zhejiang Shengdi Technology Inc.

Persona de Contacto: Mrs. Yeli Lu

Teléfono: 86-573-86123368

Fax: 86-573-86161698

Envíe su pregunta directamente a nosotros (0 / 3000)

Otros productos
Single Phase Energy Meter Test Bench
Tres Energía medidor de fase Banco de Pruebas
Portable Energy Meter Test Equipment
Solicitar una cotización

E-Mail | Sitemap

Privacy Policy China Bueno Calidad Single Phase Energy Meter Test Bench Proveedor. Copyright © 2016 - 2019 metertesting.biz. All Rights Reserved. Developed by ECER